Specifications
Features of Semi-Auto Chemistry Analyzer NEUBA22:
1.Analytical Modes: kinetic, fixed time, end-point, absorbance
Specification of Semi-Auto Chemistry Analyzer NEUBA22:
Assay Method: reaction rate, endpoint, two-point, double wavelength, coefficient, multiple standards, blank reagent, blank sample, absorbance at fixed point, Q.C, Q.C graph and turbidity Temperature System: PELTIER temperature control component, 25℃,30℃,37℃,accuracy: ±1℃ Light Source: 6V, 10W halogen lamp of more than 2000 hours’ life time GML Precision:±2nm FWHM: ≤10nm Stray Light:≤0.5%(absorbance≥2.5) Absorbance Range:-0.214~3.000 Absorbance Linearity:±1% Absorbance Repeatability:±0.001 Precision: ≤1% Carry Over: ≤1% Resolution: 1m Abs Stability: ≤3mAbs/h Flow Cell: permanent quartz flow cell, minimal volume is 32uL,and disposable flow cell is selectable Sampling System: pressure auto releasing pumps ensure the accuracy of suction volume Absorption Volume:200uL~3000uL Sample Volume:10uL~50uL Operational System: Windows XP with 1GB memory capacity Display:10.1’LCD Print: print standard test A4 test report Interface:RS-232 serial port Working Condition:Temperature:10℃-30℃,Relative Humidity:20%-80% Voltage: AC 85~264V,auto adapt according to the input voltage Power: ≤100W Quality System: CE
Be the first to review “Semi-auto Chemistry Analyzer”