Specifications
Features of Urine Analyzer
1.High speed, only 7 seconds per test: 2.LCD display, operating easy with friendly interface; 3.Gather the used strips automatically: 4.High brightness and long live light source(LED): 5.New type light sensor, high-sensitivity, good anti-interference; 6.Inbuilt thermal printer 7.Touch panel 8.RS-232 Interface for data exchange and system software update; 9.Parallel port for connecting with external printer: 10.Store 2000 samples; 11.The sensitivity of each parameter can be adjusted through software interface;
Specifications:
Parameters: GLU. BL, SG, PH. KET, BLD, PRO, URO, NIT. LEu, VC Wave length: 470nm. 540nm.625nm Principle: Reflectance photometry Strips:10 parameters.11 parameters or more Test speed:514TEST/per hour LCD model:320×240 Dot matrix Storage: 2000 test reports Printing: Built in thermal printer, or external printer as an option Working Temp:10-30℃ Relative Humidity:s70% Power Supply:AC100-240V, 50/60Hz Power:<45VA Dimensions: 390×330×210mm Weight:7kg
Specifications
|
|
Testing Principles
|
The reflective photometer reads the color on the strip to calculate the concentration of each testing item and output a semi- quantitative analysis result.
|
Testing Wave Length
|
470nm、525nm、625nm
|
Testing Speed
|
500 samples / hour
|
Stray light
|
Relative intensity ≤ 1%
|
Linearity
|
Linear correlation coefficient ≥0.995
|
Repeatability
|
Coefficient of variation CV ≤ 1%
|
Stability
|
During 8 hours of operation, Coefficient of variation CV ≤ 1%
|
Output Options
|
LCD display, serial port output, Embedded 57mm thermal printer (English/Chinese), External parallel dot matrix printer,(English/Chinese)
|
LCD Screen
|
240×64 dot-matrix LCD
|
Report Storage
|
Stores up to 2000 sets of results.
|
External Ports
|
RS-232C connected with PC 25 pin centronics port which can driver a dot external printer.
|
Working Voltage
|
AC100-240V 50/60Hz
|
Power
|
≤ 45 w
|
Dimension
|
390X330X210 mm
|
Weight
|
6 kg
|